Technical Specifications

SpO2 Specifications

FunctionMasimo M-LNCS DCIChipOx
Saturation Normal (no motion) 70-100%Arms ± 2% 70-100%Arms ± 2%
Saturation Motion 70-100%Arms ± 3% N/A
Low Perfusion SpO2 +/-2% N/A
Pulse Heart rate ± 3% BPM 20-300 BPM: ±3 BPM
Pulse Heart rate w/motion ± 3% BPM 20-300 BPM: ±3 BPM

  

Temperature Specifications

FunctionCovidien Electronic Thermometer
Temperature Accuracy Quick Mode (Oral): ± 0.3°C
Temperature Accuracy Std. Mode (Aux/Rectal): ± 0.1°C
Temperature Accuracy Direct Mode ± 0.1°C
Response Time 4 sec (quick mode); 10-15 sec. (Axillary/Rectal); 60 sec.(direct mode)
Resolution 0.1°C

  

Blood Pressure Specifications

SpecificationsValues
Range:
  • Pressure: Diastolic: 20-200 mmHg (adult), 20-160 mmHg (pediatric).
  • Systolic: 40-260 mmHg (adult), 40-230 mmHg (pediatric)
  • Heart Rate: 30-220 BPM (beats per minute)
Accuracy (BP): Meets or exceeds ANSI/AAMI/ISO 81060-2:2013 standard for non-invasive accuracy (±5mmHg mean error with 8mmHg standard deviation).

  

General Specifications

SpecificationsValues
Conditions for Use: Operating: 10°C (50°F) to 40°C (104°F) 15 – 90% RH non-condensing - 700 kPa - 1060 kPa. Operating the monitor in an environment at maximum temperature can produce temperatures exceeding 41°C (41.6°C highest recorded) on a patient-applied part. It is up to the operator to determine if this temperature is too high based upon the condition of the patient and, if so, to ensure the ambient temperature of the environment is 38°C or below.
Storage: -20°C (-4°F) to 55°C (149°F) 15 – 90% RH non-condensing - 500 kPa - 1060 kPa. Performance can be affected if used or stored outside the specified temperature, humidity, or altitude listed in the ranges above.
Power: Internal power supply. Input: 100-240 VAC @ 1.5A max, 50-60 Hz. Output +9VDC @ 5A IEC 320 type input connector.
Battery: Type: 7.2v, 6600mAh Internal Lithium-Ion battery Run Time (full charge): ~20 hrs* / 200 BP cycles Charge Time: ~6 hrs @ 23deg Celsius +/- 2deg *Unit with BP, Temp, SpO2, Wi-Fi, barcode scanner
Calibration: The accuracy of cuff -pressure transducers/indicators should be verified bi-annually.
Safety Systems: Independent hardware over-pressure circuit and redundant software overpressure algorithm to limit cuff pressure to less than 300 mmHg (+20/-10mmHg). Independent hardware timing circuit and redundant software timer algorithm to limit the duration of a blood pressure cycle to less than 180 seconds.
Dimensions: Size (without thermometer option): 8” H x 6.75’’ W x 5.2’’ D or 215mm H x 190mm W x 140mm D
Standards: ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012, EN 60601-1:2006/A1:2013, IEC 60601-1-2: 2007 EMC, IEC 80601-2-30: 2013, ISO 80601-2-61: 2011, ISO 15223-1:2012, ISO 10993-1, 2009, ISO 10993-5, 2009, EN ISO 10993-10, 2010, ISO 81060-2:2013, EN 50419: 2006, EN ISO 14971:2009, CSA C22.2 No. 60601-1, EN ISO 81060-1: 2012, IEC 60601-1-6: 2013
Classifications: Equipment Classification: Class IIa per MDD, Class I (Electrical Shock), Continuous mode of operation, CE
Ingress of liquid: Ingress Protection: IPX1: This device is protected against harmful effects of dripping water per IEC 60529.

  

Limited Warranty

SunTech Medical, Inc. provides to the original purchaser the following limited warranty from date of invoice:

SpecificationsValues
Main unit, SpO2 modules, temp modules: 3 years
OPD Blood Pressure Cuffs: 2 years
Accessories (SpO2 sensors, temp probes, BP hose, etc): 90 days

Masimo® and SET® are registered trademarks of Masimo Corporation. Covidien® and FILAC® are registered trademarks of the Covidien AG Corporation. Nellcor® is a registered trademark of Covidien LP Limited Partnership.

   

Parts & Accessories

Base Unit Contents

  • SunTech CT40 Device with BP
  • AC Power Cord
  • 10’ BP Patient Hose
  • Battery
  • Quick Start Guide
  • Adult Plus OPD Cuff
  • Large Adult Plus OPD Cuff

   

SunTech® CT40 System & Modules

DescriptionPart Numbers
SunTech CT40 Base unit with BP (no power cord) 99-0134-00
SunTech CT40 Base unit with BP with AC Power Cord (Americas) 99-0134-01
SunTech CT40 Base unit with BP with AC Power Cord (Europe) 99-0134-02
SunTech CT40 Base unit with BP with AC Power Cord (UK) 99-0134-03
Chipox Nellcor® Compatible SpO2 Module 98-0411-00
Masimo SET® SpO2 Module (preferred) 98-0411-01
Covidien FILAC® 3000 Thermometry Module 98-0411-02

   

Systems Reference Configuration

DescriptionPart Numbers
BP Only 99-0134-0X
BP, Masimo® SpO2 99-0134-0X + 98-0411-01
BP, Nellcor® comp. SpO2 99-0134-0X + 98-0411-00
BP, FILAC® Temp 99-0134-0X + 98-0411-02
BP, FILAC® Temp, Masimo® SpO2 99-0134-0X + 98-0411-02 + 98-0411-01
BP, FILAC® Temp, Nellcor® comp. SpO2 98-0134-0X + 98-0411-02 + 98-0411-00

   

SpO2

   

Masimo SET® SpO2 Module Accessories

DescriptionPart Number
Masimo® M-LNCS DCI Adult SpO2 Reusable Sensor 2501*
Masimo® M-LNC 10’/3.0m Patient Cable 2525*
*Contact Masimo® for replacement orders

   

ChipOx Nellcor Compatible SpO2 Module

DescriptionPart Number
Reusable Adult SpO2 Finger Sensor, ChipOx 52-0010-00

   

Temperature

   

Covidien FILAC 3000 Temperature Module Accessories

DescriptionPart Number
F3000 Oral/Axillary Temp Probe, 9’ 52-0009-00
F3000 Oral/Axillary Isolation Chamber, Blue 45-0006-00
F3000 Rectal Temp Probe, 9’ 52-0009-01
F3000 Rectal Isolation Chamber, Red 45-0006-01
F3000 Disposable Temp Probe Covers (Qty 500 - 25 boxes of 20) 98-0131-00
F3000 Disposable Temp Probe Covers (Qty 5000) 98-0130-00
F3000 Calibration Plug 52-0011-00

   

Accessories

 

DescriptionPart Number
Deluxe Mobile Stand 46-0040-00
SunTech CT40 Barcode Scanner (w/USB) with Scanner Mount 98-0257-00
SunTech CT40/50 WiFi Dual Band USB Dongle 45-0005-00
10’/3.0m Patient BP Hose, Bayonet to Bayonet 91-0028-16
AC Power Cord (Americas) 91-0003-00
AC Power Cord (Europe) 91-0003-05
AC Power Cord (UK) 91-0003-06
Mini-USB to USB Cable 91-0100-02
SunTech CT40 Rechargeable Lithium Ion Battery 17-0022-00
SunTech CT40 1D Barcode Scanner 98-0257-03
SunTech CT40 2D Barcode Scanner 98-0257-04

   

Blood Pressure Cuffs

 

Blood Pressure Cuffs (Box of 5)

DescriptionPart Number
OPD Child (12-19cm) BP Cuff, Bayonet 98-0600-41 
OPD Small Adult (17-25cm) BP Cuff, Bayonet 98-0600-43 
OPD Adult (23-33cm) BP Cuff, Bayonet 98-0600-45 
OPD Adult Plus (28-40cm) BP Cuff, Bayonet 98-0600-4A
OPD Large Adult (31-40cm) BP Cuff, Bayonet 98-0600-47
OPD Large Adult Plus (40-55cm) BP Cuff, Bayonet 98-0600-4C

         

Looking for a complete list of Parts & Specs for each product? Click here.

 

BP+ Vitals Devices

Looking for more information about on our BP+ Vitals? Click on the links below to find out more.

We use cookies
Cookies are small text files that are downloaded onto your device when accessing this website and may entail sharing of data with 3rd parties. SunTech's cookies can be categorized as Required (necessary for the website to function properly, such as providing secure log-in or remember your preferences), Analytical (collect statistical usage data for website optimization) and/or Targeted Advertising (track your activity to deliver marketing content tailored to your interests, within and beyond SunTech websites). Click “ACCEPT” to agree to our terms and allow cookies or click “DENY" to refuse. If you choose "DENY" you may still access portions of this site, however, some functionality will not work.