Why is it important to use the correct size cuff on my patients?

SunTech Medical Cuffs are designed to meet the AAMI (Association for the Advancement of Medical Instrumentation) clinical standards for blood pressure cuffs. These standards were devised in order to ensure adequate pressure on the artery of the upper arm, which is necessary to produce an accurate blood pressure measurement. If a cuff is used on an arm larger than the specified range, the resulting reading will be falsely high. If a cuff is used on an arm that is smaller than the specified range, the resulting reading will be falsely low

You can find out more by clicking here for a useful video tutorial.

How do I clean the Orbit-K cuff after a stress test?

Please refer to the instructions for use.

How do I clean the K-sound Microphone?

The K-sound Microphone can be cleaned using a mild disinfectant wipe (e.g. Sani-Cloth®) and letting the microphone line dry.

You can find out more by clicking here for a useful video tutorial.

Will my Tango+ K-sound Microphone work with my Tango M2?

No, the connector of the K-sound microphone used with the Tango+ is different than the connector of the K-sound microphone used with the Tango M2. Please see the Technical Specs Tab to view which Orbit-K cuffs will work with which stress BP monitor.

end faq