Advantage A+

Advantage A+

The highest performance OEM module in the smallest complete package available. SunTech designed the A+ to meet the needs of the most challenging applications. With internal automatic modes, low voltage communication protocols and the lowest power consumption in the industry, A+ provides the solution for the most demanding needs.

Technical Specifications

Physical Dimensions
Width2.0 in [50.8 mm]
Height3.5 in [88.9 mm]
Length1.0 in [25.4 mm]
Weight0.25 lbs [113.4 g]
Measurement
TechniqueOscillometric
NIBP AccuracyMeets ANSI/AAMI SP10-2002, EN 1060-4
Patient ApplicationsAdult, Pediatric, Neonatal
Systolic Range
Adult40-260 mmHg
Pediatric40-160 mmHg
Neonatal40-130 mmHg
Diastolic Range
Adult20-200 mmHg
Pediatric20-120 mmHg
Neonatal20-100 mmHg
Pulse Rate Range30-220 BPM (Beats per min)
Pulse Rate Accuracy±2% or ±3 BPM, whichever is greater
Transducer Accuracy± 3 mmHg over full range in operating conditions
Operating ModesManual, Long-term automatic, Short-term automate (STAT), Service
Operating Conditions0˚C to 50˚C, 15% to 95%, non-condensing humidity
Storage Conditions-20˚C to +65˚C, up to 90% RH non-condensing
Electrical
Operating Voltage5.3-14.5 VDC (High), 1.8-9.0 VDC (Low)
Power ConsumptionUsing 6 VDC Nominal (High) or 3.3 VDC Nominal (Low)
Sleep2.0 mW (High), 3.0 mW (Low)
Idle372 mW (High), 430 mW (Low)
Measurement1.3 W (High), 1.4 W (Low)
Inflation3.7 W (High), 3.4 W (Low)
Communication ProtocolsSerial RS232, TTL (1.8, 3.3, or 5V), or I 2 C ready
Safety & Regulatory StandardsIEC 60601-1:1997 IEC/EN60601-2-30: 1999/2000 AAMI SP10:2002, EN1060-1:1996 EN1060-3:1997, EN1060-4: 2004
Advantage Technologies
SMT (Standard Motion Tolerance)
TMT (Transport Motion Tolerance)
HDM (Hemodialysis Monitoring)
RMT (R-wave Motion Tolerance)
VET (Veterinary Monitoring)
Customer Support
Standard Support
Custom Support10 hours*
Custom DesignAdditional Charge
Warranty
Time2 years
Cycles30,000
*additional support charged at standard rates