Advantage Series of Oscillometric OEM NIBP Technologies
Overview
The Advantage series of oscillometric OEM NIBP technologies from SunTech Medical provides the simplicity of the oscillometric technique of automated blood pressure with the highest level of clinical performance and flexibility in the industry.
Since every OEM application presents a unique set of challenges, the Advantage series offers multiple application specific technologies designed for specific clinical environments and patient populations. There are currently five Advantage NIBP technologies available to choose from to meet your exact requirements.
Advantage SMT – Standard Motion Tolerance
SunTech Medical's standard motion tolerant technology has been designed to meet the needs of most clinical monitoring applications. Currently used in many bedside, multi-parameter and ambulatory monitors, delivering clinical accuracy and patient motion tolerance.
Advantage HDM – Hemodialysis Monitoring
Designed to provide accurate and reliable blood pressure monitoring during hemodialysis treatment by researching the complexities of changing relative blood volumes and compromised physiologies of end-stage renal disease patients
Advantage TMT – Transport Motion Tolerance
SunTech Medical worked with ambulance and helicopter EMS services to develop an automated NIBP technology designed to perform in the presence of motion and noise artifact encountered during ground and air transport conditions.
Advantage RMT - R-wave gated Motion Tolerance
SunTech Medical’s most advanced motion tolerant technology uses the same principle of R-wave gating (ECG gating) utilized on the Tango+™ treadmill stress and exercise monitor for the most challenging patient motion environments.
Advantage VET – Veterinary Monitoring
SunTech developed the VET technology in co-operation with a leading veterinary medical school to provide the most accurate BP results in both small and large animals.
A. No. NIBP simulator target values should never be considered as a standard of accuracy. All NIBP simulator manuals explain or should explain that target values are approximations and should only be used for measuring repeatability. Accuracy can only be measured using human subjects as described in the accuracy requirements and test methods given in the AAMI/SP10 Standard for Automated Sphygmomanometers. All SunTech Medical NIBP algorithms are tuned specifically for use on live human or animal subjects, not NIBP simulators. Please refer to the white paper “NIBP Simulator Limits with the Advantage OEM NIBP Module Series” for more details on how simulators are to be used. Included in that paper are a description of a correct setup and a comparison of five common NIBP simulators. Further information on interaction with simulators is also available in Appendix B in the Advantage 2.0 and Mini Operators Manual.
A. The Advantage OEM Blood Pressure modules are field programmable. Additional features and improvements can now be uploaded to the module via the RS232 or TTL serial communication interface without any jumpers, special cables or complicated instructions. This saves time and money since modules no longer need to be returned to the factor every time new features or improvements become available.
A. The pump is never on while collecting data. Thus any noise generated by the pump, regardless of the model, cannot contaminate or alter the oscillometric data used to build the curve.
A. While there are small differences between the devices, the Mini was designed to be a functionally equivalent but physically smaller version of the 2.0, and provides the option of mounting the pump and valves separately from the electronics. All functional parts of the hardware relating to the acquisition of BP are identical. Thus with any specific version of the firmware code, operation of the two units will be identical. The Advantage 2 has RS232 and TTL serial communication interface options. The Advantage Mini has no RS232 level communications option in its original configuration, therefore it lacks the RS232 level translator chip and ancillary components. Please refer to the operator's manual for complete description, instructions, warnings, cautions and complete specifications.
A. The Advantage series of OEM Non-Invasive Blood Pressure modules does meet AAMI SP10 -2002 specifications. They also meet the EN60601-1, EN60601-2-30, EN1060-1 and EN1060-3 standards.
A. Both processors have separate connections. If the safety microprocessor detects either an overpressure or timeout condition, it will immediately override the BP microprocessor and and cut power to the pump and valves thus deflating the system.